Short description
1.1 This test method covers an immunological method to determine the amount of antigenic protein in Hevea Natural Rubber and its products using rabbit antisera specific for HNRL proteins. This immunoassay procedure quantitatively measures the level of antigenic latex proteins in solution using an inhibition format. The samples may include glove or other product extracts which have been collected in order to measure the HNR levels. Although this method detects antigenic proteins, it should not be considered as a measure of allergenic proteins. Correlation of protein/antigen levels with the level of allergenic proteins has not been fully established. 1.2 For the purpose of this test method, the range of protein will be measured in terms of microgram to milligram quantities. 1.3 This test method can be used to determine the presence and amount of HNR protein contamination on test items. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.