Short description
1.1 This test method covers the precise on-line determination of pH in water samples of conductivity lower than 100 µS/cm (see Table 1 and Table 2 ) over the pH range of 3 to 11 (see Fig. 1 ), under field operating conditions. pH measurements of water of low conductivity are problematic for conventional pH electrodes, methods, and related measurement apparatus. TABLE 1 Calculated Conductivity and pH Values at 25°C of Low Concentrations of NaOH in Pure Water A Note 1 - This table tabulates the theoretical conductivity and pH values of low levels of NaOH in pure water as calculated from available thermodynamic data. Note 2 - To illustrate the high sensitivity of the sample pH at these low concentrations to contaminants, the last column lists errors that would result if the sample were contaminated with an additional 1 mg/L through sample or equipment handling errors. [...] A Data courtesy of Ref (1) . B This data developed from algorithms originally published in Ref (2) . B The boldface numbers in parentheses refer to a list of references at the end of this standard. TABLE 2 Calculated Conductivity and pH Values at 25°C of Low Concentrations of HCl in Pure Water A Note 1 - This table tabulates the theoretical conductivity and pH values of low levels of HCl in pure water as calculated from available thermodynamic data Note 2 - To illustrate the high sensitivity of the sample pH at these low concentrations to contaminants, the last column lists errors that would result if the sample were contaminated with an additional 1 mg/L through sample or equipment handling errors.[...] A Data courtesy of Ref (1) . This data developed from algorithms originally published in Ref (2) . FIG. 1 Restrictions Imposed by the Conductivity pH Relationship 1.2 This test method includes the procedures and equipment required for the continuous pH measurement of low conductivity water sample streams including the requirements for the control of sample stream pressure, flow rate, and temperature. For off-line pH measurements in low conductivity samples, refer to Test Method D5464. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.